MIRM ROOM PRESSURE MONITOR About
| Display | High light 3.2 color TFT LCD, resolution 320 240 |
| Display Panel Material | PMMA |
| Resolution (Display Unit) | 0.1 engineering unit |
| Channels | 13 channels (parallel or alternate display) |
| Engineering Units | C / F, %RH, Pa |
| Update Time | < 1 second |
| Front Panel Material | 316L stainless steel (1.5 mm thick) |
| Front Panel IP Rating | IP65 |
| Sensor Cap IP Rating | IP54 |
| Back Housing Material | Fire-proof ABS + PC (UL94 V-0) |
| Weight | ~450 g (horizontal) |
| Power Supply | 24 VDC |
| Sensor Range | Temperature: 050C; Humidity: 0100% RH; Differential Pressure: 060 Pa |
| Accuracy | Temperature: 0.30.5C; Humidity: 2080% RH; Diff. Pressure: 0.51% FS |
| Non-Linearity | Humidity: <0.1% RH |
| Repeatability | Temperature: 0.1C; Humidity: 0.1% RH |
| Hysteresis | Humidity: 1.0% RH |
| Long-Term Drift | Temperature: <0.02C/year; Humidity: <0.25% RH/year; Diff. Pressure: <0.5% FS/year |
| Response Time | Temperature: <120 s; Humidity: <40 s; Diff. Pressure: 0.530 s |
| Temperature Drift | Diff. Pressure: <0.05% FS/C (zero), <0.08% FS/C (span) |
| Temperature Compensation | Diff. Pressure: 050C |
| Medium Temperature | Diff. Pressure: 060C |
| Overload Capacity | Diff. Pressure: 10 FS (overload), 15 FS (burst) |
| Analog Inputs | Max. 3 (420 mA / 010 V), accuracy <0.1% FS |
| Input Range | Default: 050C / 0100% RH / 060 Pa; Optional: -50100C / 0100% RH / -100100 Pa |
| Analog Outputs | Max. 3 (010 V), accuracy up to 0.2% FS, load >2 k |
| Keys / Functions | Alarm set/reset, DP re-zero, calibration, display mode setting |
| Communication | USB (parameter setup), RS485 Modbus RTU (9600 baud) |
| Terminal Size | Max. 1.5 mm |
| Working Environment | 050C, 095% RH (non-condensing) |
| Storage Environment | -10C to 70C |
| Process Connections | Built-in T/RH sensor (front filter cap); Diff. pressure: rear nozzles (5 mm) or front screw |
| Certification | Calibration certification |
ABOUT
Our room pressure monitor provides real-time monitoring of differential pressure within enclosed spaces. It ensures optimal air quality and safety by detecting and maintaining desired pressure differentials, crucial in cleanrooms, hospitals, laboratories, and pharmaceutical facilities. With its precise measurements and configurable alarm thresholds, it safeguards against contamination and ensures compliance with regulatory standards
Features
- Real-time room pressure monitoring
- Maintains optimal air quality and safety
- Essential for cleanrooms, hospitals, labs
- Configurable alarms for compliance
- It gauges temperature and humidity alongside differential pressure
FAQ
Q: What is the MIAL Room Pressure Monitor used for?
A: It is used to monitor and control room pressure in critical environments.
Q: Where is the MIAL Room Pressure Monitor commonly used?
A: It is commonly used in hospitals, clean rooms, laboratories, and pharmaceutical facilities.
Q: Can the room pressure monitor display differential pressure values?
A: Yes, it provides accurate differential pressure indication and monitoring.
Q: Does the MIAL Room Pressure Monitor support alarm functions?
A: Yes, it can provide audio or visual alarms for pressure deviations.
Q: Is the MIAL Room Pressure Monitor suitable for HVAC applications?
A: Yes, it is ideal for HVAC and clean room pressure monitoring applications.