Call us
07315134316
Feature
| Operation and performance | |
|---|---|
| Pipe sizes | 15 MM -2000 MM |
| Conductivity | 20 S/cm |
| Pipe materials | Metallic and Non Metallic pipes. |
| Flow accuracy | Standard :0.5% Optional: 0.2% Achievable with process calibration |
| Repeatability | Flow:0.15% BTU: 0.27% |
| Linearity | Standard: 0.5% Optional: 0.2% |
| Meassuring range | 0.09ft/s - 40 ft/s |
| Measurement parameters | Flow Meter- Instantaneous flow rate, totalized flow Btu meter - Instantaneous energy rate, totalized energy, Instantaneous flow rate, totalized flow,supply temperature and return temperature |
| Certification | Factory calibration certification, CE, ISO |
| Electronics | |
| Enclosures ABS | Use weather proof enclosure while installing the transmitter outside |
| Enclosure IP rating | IP 65 |
| Memory | EEPROM |
| Power supply | 24 VDC/2A Use 2-amp SMPS when employing AC power |
| Ambient temperature | 32F to 140F ( 0 to 60) |
| Relative Humidity | 5- 95% RH |
| Standard Analog outputs | Flow meter- 4-20 mA Output programmed for current flow rate. 750 maximum load, Btu meter- 4-20 mA output programmed for current flow rate or current energy rate. 750 maximum load, |
| Pulse Outputs | Flow Meter- Pulse Programmed for Flow Consumption , Contact pulse Duration -0.1~300 ms Btu meter - Pulse Programmed for Enery Consumption or Flow consumption , Contact pulse Duration -0.1~300 ms |
| Network Connection | Modbus RTU RS485 |
| Cable | 10 m |
| Flow tube specification | |
| Coil material | Pure Copper 99% copper (Cu) content, excellent electrical conductivity, corrosion resistance |
| Process connections | Standard: ANSI 150 flanges Optional: ANSI 300 flanges |
| Operating temperature | 14F to 248F (-10C to 120C) |
| Nominal Pressure | Standard: 1.6 Mpa Optional: 2.5 Mpa, 4.0 Mpa,5Mpa |
| Flow Tube | Standard :SS 304 Optional : SS 316 Electrode Material Standard: SS 316L Optional: Hastelloy,Titanium, Tantalum, |
| Liner | Standard: PTFE Optional: Ebonite, Polyebonite, Polyurethane,PFA |
| Flange | Standard: Carbon Steel Optional: Stainles steel |
| Mountings | Flanged flow tube |
| IP rating | Flow tube : IP68 Tramsmitter : IP 65 |
| Energy measurement | |
| Temperature sensor PT1000 | -22F to 392F (-30 C to 200 C) |
| Wetted Immersion Temperature sensor with Thermowell | |
| Cable | 10 M |
Step into the world of precision measurement with the kingly MEF 2100 Inline Electromagnetic Flow Meter by Mial Instruments. Shop Now for the latest in flow technology designed for laureate performance in water monitoring. Polished with excellence, this flow meter features ANSI 150 flanges, a digital LCD display, and standard analog output (4-20 mA). With unrivaled accuracy of 0.5%, it suits diameters from 15mm to 2000mm and offers reliable operation from -10C to 120C. Benefit from IP65/IP68 protection, robust SS or MS construction, and clearance options for exporters, manufacturers, and suppliers across India.
MEF 2100 Application and Usage
The MEF 2100 Inline Electromagnetic Flow Meter is primarily used for accurate water flow measurement in various industrial and municipal applications. Direction of use involves inline installation, aligning with process pipelines for continuous monitoring. It caters specifically to water as the application media, ensuring reliable data collection whether for treatment plants, distribution systems, or process industries. With a user-friendly digital LCD display, operators can easily interpret real-time results, making flow management both efficient and precise.
Certifications, Payments, and Sample Policy
The MEF 2100 Flow Meter is supplied with all necessary certifications to meet international standards. Payment terms are flexible to facilitate smooth business, while we ensure goods transport is swift and secure with stock ready for handover upon order confirmation. For new clients, our sample policy supports preliminary quality checks, allowing you to experience the product's reliability before extensive procurement. Trust seamless processing from order to delivery, ensuring your operational needs are met without delay.

Price:
Minimum Order Quantity : 1 Unit
Stability : Long term: <0.1% reading/year
Resolution : 0.01 m3/h
Temperature Range : 10C to 60C
Power Supply : 230V AC, 50 Hz
Flow Rate : 0.03 6000 m3/h
Minimum Order Quantity : 1 Unit
Stability : Flow: 0.15%; BTU: 0.27%
Resolution : Analog Output: 420 mA
Temperature Range : 22F to 392F (30C to 200C)
Power Supply : 24 VDC / 2A (Use 2A SMPS for AC power)
Flow Rate : Corresponds to measuring range